µQuant™ Microplate Spectrophotometer
|
| Detection method | Absorbance |
| Read method | Endpoint and spectral scanning. Kinetic and well area scanning via computer control. |
| Microplate types | 6-, 12-, 24-, 48-, 96-, 384-well microplates and 60-/72-/96-well Terasaki plates. Also reads up to 8 Bio-Cells (patented quartz miniature cuvettes) for 1 cm absorbance measurement. |
| Reading speed | 96=45 sec |
| Wavelength range | 200 to 999 nm |
| Wavelength selection | Monochromator, selectable 1 nm or greater increments |
| Monochromator wavelength accuracy | + 2 nm |
| Monochromator wavelength repeatability | + 0.2 nm |
| Bandpass | 2.4 nm |
| Dynamic range | 0 to 4.0 OD typical |
| OD accuracy | <1% + 0.01 OD typical |
| OD linearity | +1% typical |
| OD repeatability | 0.5% + 0.005 OD typical |
| Light source | Xenon flash lamp |
| Stray light | 0.03% at 230 nm typical |
| Onboard software | Multiple pre-programmed assays including DNA quantitation and 260/280 ratio |
| Dimensions | 16" D x 15" W x 7" H (41 x 38 x 18 cm) |
| Weight | 25 lbs (11.34 kg) |
| Regulatory | For in-vitro diagnostic use (Except in the European Union). All BioTek microplate instrumentation is CE and ETL marked. For details on approvals and standards compliance, please contact BioTek. |
Specifications subject to change.