Synergy™ HT Multi-Detection Microplate Reader
BioTek

Synergy™ HT Multi-Detection Microplate Reader Synergy™ HT Multi-Detection Microplate Reader
bullet off Overview
bullet off Features
bullet off Models
bullet off Specifications
bullet off Accessories
bullet off Print Product Details
bullet on Specifications

Detection method Fluorescence, Time Resolved Fluorescence, Luminescence: Photomultiplier tube
Absorbance: Silicon photodiode
Read method Fluorescence and Luminescence: Endpoint, kinetic, well area scanning
Absorbance: Endpoint, kinetic, well area scanning, spectral scanning
Microplate types 6- to 384-well plates
PCR tube trays
Terasaki trays
Reading speed Fluorescence and Luminescence: 35 sec/96 wells (depends on settings)
Absorbance: 15 sec/96 wells (depends on settings)
Wavelength range Fluorescence: Excitation 300-650 nm (200-700 nm with TR model),
Emission 300-700 nm (red extended option available)
Time Resolved Fluorescence: Excitation 200-700 nm, Emission 300-700 nm (red extended option available)
Luminescence: 300-700 nm (red extended option available)
Absorbance: 200 to 999 nm, tunable in 1 nm increments
Wavelength selection Fluorescence: Deep blocking interference filters
Time Resolved Fluorescence: Monochromator
Absorbance: Monochromator
Monochromator wavelength accuracy +/- 2 nm
Monochromator wavelength repeatability +/- 0.2 nm
Filter wheel capacity Fluorescence: 4 excitation positions, 4 emission positions
Absorbance: N/A
Filters supplied Fluorescence: 4 excitation filters, 4 emission filters
Absorbance: N/A
Bandpass Fluorescence: Variable, dependent on filter
Time Resolved Fluorescence: 10 nm
Absorbance: 2.4 nm
Dynamic range Fluorescence: 0 - 99,999 RFU's
Time Resolved Fluorescence: 0 - 99,999 RFU's
Luminescence: 0-99,999 RLU's
Absorbance: 0 - 4.0 OD
OD accuracy < 1% at 2.0 OD typical
< 3% at 3.0 OD typical
OD linearity < 1% from 0 to 3.0 OD typical
OD repeatability < 0.5% at 2.0 OD typical
Resolution 0.001 OD
Sensitivity Sodium fluorescein 2.5 pg/ml (1.3 fmol/well) typical*
Sodium fluorescein 1 pg/ml (0.5 fmol/well) limit of detection*
Luciferase 0.1 pg/well typical
ATP 100 amol/well limit of detection
(*Black plate, top measurement, 3mm diameter probe)
Standard probes 3 mm top, 5 mm bottom. Top probe height and top or bottom reading position are software adjustable/selectable (fluorescence). Other probe sizes available.
Light source Fluorescence: Tungsten Halogen 1,000 hrs lifetime
Time Resolved Fluorescence: Xenon flash lamp, 1 billion flashes lifetime
Absorbance: Xenon flash lamp, 1 billion flashes lifetime
Stray light Absorbance: 0.03% at 230 nm typical
Temperature control 4oC above ambient to 50oC
Shaking Linear shaking: Low, med, high and variable speeds
Reagent dispensers 2 syringe pumps
Dispense volume 5-1000 µl in 1 µl increments
PC Software Gen5 software.
Control through USB or serial port.
Power 100-240 VAC 50/60 Hz
Dimensions 16"W x 15"D x 10"H
(40.6 x 38 x 25.4 cm)
Weight 40 lbs (18 kg)
Regulatory For in vitro diagnostic use (Except in the European Union).
All BioTek microplate instrumentation is CE and ETL marked. For details on approvals and standards compliance, please contact BioTek.
Automation Compatible with Bio-Stack Microplate Stacker and Bio-Stack Twister II Microplate Handler.

Specifications subject to change.

 