PowerWave™ Microplate Spectrophotometer
|
|
| Detection method | Absorbance |
| Read method | Endpoint, kinetic, spectral, linear and well area scanning (depending on model) |
| Microplate types | 96- and 384-well microplates (PowerWave and PowerWave HT models); 6- to 384-well microplates and 60-/72-/96-well Terasaki plates (PowerWave XS model); Also reads up to 8 Bio-Cells for 1 cm measurement. |
| Reading speed | PowerWave and PowerWave HT: PowerWave XS: |
| Wavelength range | 200 to 999 nm; 340 to 999 nm (PowerWave HT 340 and PowerWave 340 models) |
| Wavelength selection | Monochromator, selectable 1 nm or greater increments |
| Monochromator wavelength accuracy | +2 nm |
| Monochromator wavelength repeatability | +0.2 nm |
| Bandpass | 5 nm (PowerWave and PowerWave HT) 2.4 nm (PowerWave XS) |
| Dynamic range | 0 to 4.0 OD |
| OD accuracy | 1% + 0.01 OD typical |
| OD linearity | +1% typical |
| OD repeatability | 0.5% + 0.005 OD typical |
| Resolution | 0.001 OD |
| Light source | Xenon flash lamp |
| Stray light | 0.03% at 230 nm typical |
| Temperature control | Ambient +4oC to 50oC 4-Zone™ natural convection |
| Shaking | User-programmable timing and frequency |
| PC Software | Includes Gen5 or KCjunior software (depending on model) |
| Power | 100 VA max; Instrument powered by universal input (100-240 VAC 50-60 Hz) power supply. |
| Dimensions | 16" D x 8.5" W x 8.5" H (40.6 x 21.6 x 21.6 cm) |
| Weight | 24 lbs (11 kg) |
| Regulatory | For in-vitro diagnostic use (Except in the European Union). All BioTek microplate instrumentation is CE and ETL marked. For details on approvals and standards compliance, please contact BioTek. |
| Automation | Compatible with Bio-Stack Microplate Stacker and Bio-Stack Twister II Microplate Handler. |
Specifications subject to change.